Aeromedical oxygen calculator

HAMILTON-T1 · Transport oxygen planning · New Zealand cylinder sizes

Patient support

Respiratory support
Fraction 0.21–1.00
fraction
L/min, from T1 ExpMinVol
L/min

Use measured expiratory minute volume (ExpMinVol), not the oxygen flowmeter setting.


Transport mode

Estimated safe transfer time

1 × D cylinder · 1,490 L nominal
—minutes
Estimated remaining transfer time after oxygen and contingency allowances.
Loading estimates…
Estimated T1 O₂ consumption
—
L/min · manufacturer formula
Allowable planning demand
—
L/min · includes +25%
Usable selected-cylinder O₂
—
L · 25% retained
Estimated time before buffer
—
minutes · after 25% gas reserve

Oxygen cylinder duration

Cylinder size comparison

Supply

Selected: 1 × D, 1,490 L

CylinderCapacity1 cylinder2 cylinders
Wellington ICU · v0.3

About this calculator

Method, safety margins, cylinder information and references.

Planning estimate, not a verified safe operating limit. Confirm the actual cylinder gauges, regulator, accessible reserves and the T1's System → Info → O₂ consumption reading before departure. This calculator assumes full cylinders, a leak-free adult/large paediatric circuit, and no oxygen-driven nebuliser.

Calculation and planning assumptions

Displayed safe minutes = floor[(75% of nominal capacity ÷ 125% of estimated consumption) − 30 min], minimum 0.

Cylinder illustrations

Illustrations are scaled by the cube root of nominal oxygen content, with a common baseline (a volumetric visual comparison). They are not measured physical cylinder dimensions. The selected cylinder count is drawn individually. Select a cylinder to update the transfer time above.

Cylinder duration table

Times are in whole safe minutes, rounded down. A zero means the reserve-plus-contingency rule is not met; it does not mean the cylinder is empty. If using two cylinders, both must be accessible, compatible and reliably switchable during the transfer. E and F sizes are from an older BOC NZ product chart and require confirmation of current availability and actual labelled content.

Equipment and altitude

Altitude

Road transfer: no cabin-altitude effect. Vehicle choice changes the safety warning, not calculated oxygen consumption.

Ventilation limitations

For adults and patients above the Hamilton Method II threshold (height >70 cm and IBW >8 kg). Not valid for neonates, small paediatrics, leaking NIV circuits or an oxygen-driven nebuliser.

Calculation method and built-in safety margins

Hamilton's HAMILTON-T1 operator's manual v3.1.x, §3.4.2.2 (larger ventilated patients) and §7.5.4 (HiFlowO₂), provides these approximate oxygen supply formulas:

Ventilated: (ExpMinVol + 3) × (FiO₂ − 0.209) / 0.791 L/minHiFlowO₂: Set flow × (FiO₂ − 0.209) / 0.791 L/min

Oxygen comes from the medical O₂ cylinder and the T1 blends it with ambient air. The ventilator's dual-limb breathing circuit is not a rebreathing anaesthetic circuit. An additional altitude-dependent gas-use factor is not supported by the cited manufacturer formulas and is not assumed here.

  • 25% nominal cylinder reserve: only 75% of stated full capacity is used in the calculation.
  • 25% higher consumption: calculated oxygen demand is multiplied by 1.25.
  • 30-minute contingency: an extra 30 minutes is subtracted from the result.
  • Round down: whole safe minutes, never rounded up.

These safety margins are conservative planning choices, not manufacturer-validated clinical thresholds. They cannot cover all deterioration, leaks, diversion time, ambient pressure effects or device faults. Confirm full cylinder pressure, actual content, correct pressure regulator, delivery pressure and aircraft use restrictions before deployment.

When this calculator must not be used alone
  • Neonatal or small paediatric ventilation: Hamilton's alternative Methods I and III require different base-flow and circuit-compressibility calculations.
  • Significant leaks or NIV: oxygen use may be much higher than predicted.
  • Oxygen-driven nebuliser in the breathing circuit: added gas consumption must be accounted for separately.
  • High-altitude flight: reduced atmospheric pressure changes oxygen partial pressure and may alter clinical FiO₂ and T1 pressure delivery requirements.
  • Different T1 hardware/software versions, oxygen supplies or external concentrators: verify your local equipment setup and monitor actual O₂ use.
  • Uncertain cylinder capacity, regulator performance, unavailable cylinders, difficult switching or a high-flow rate outside the configured device's limits.

For supported T1 versions using high-pressure oxygen, the System → Info screen gives the current O₂ consumption estimate, averaged over the previous five minutes after initialisation. Use that actual reading when available, especially if it is greater than this calculated estimate.

Manufacturer and cylinder references

HAMILTON-T1 operator's manual, software 3.1.x (30 September 2024), §3.4.1, §3.4.2 and §7.5.4.
Hamilton Medical: High-flow nasal cannula therapy.
BOC Healthcare: Medical Oxygen E.P. product data sheet, NZ cylinder table (2013; verify individual cylinders).
BOC New Zealand: Medical oxygen cylinder D (current catalogue lists D and D2 separately).
BOC NZ: Cylinder size identification (actual cylinder shape and height vary by package).